Pisces

Contact Us

Technical Specifications

Display
: 12.1” TFT/LED/LCD Touch screen
Resolution : 800 x 600
Number of traces : 10 waveforms
I/O
LAN : 1 Standard RJ45 port
WLAN : IEEE 802.11b/g/n
USB : 2 USB connectors
SD : 1 SD card socket (Optional)
VGA : 1 VGA monitor connector (Optional)
Output : 1 connector for Nurse call, Defib Sync Analog Output
ECG
Lead type : 3 Lead, 5-Lead, 12-Lead
ECG waveform : 2 Channels, 7 Channels, 12 Channels
Display sensitivity : 2.5mm/mV (x0.25), 5mm/mV (x0.5) 10mm/mV (x1.0), 20mm/mV (x2.0)
Wave sweep speed : 6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth : Diagnostic mode : 0.05Hz~100Hz
Monitor mode: 0.5Hz~40Hz
Surgery mode: 1Hz~20Hz
Strong filter mode: 5Hz~20Hz
CMRR>100dB
Notch : 50/60Hz notch filter can be set to on or off Differential input impedance>5MΩ
Electrode polarization voltage range : ±400mV
Baseline recovery time<3s after defibrillation (in monitor
and surgery mode)
Calibration signal : 1mV (peak-peak), accuracy ±3%
RESP
Measurement method : Thoracic electrical bioimpedance
Measuring lead : Lead I, II
Wave gain : x0.25, x0.5, x1, x2
Respiratory impedance range : 0.5-5Ω
Baseline impedance : 500-4000Ω
Gain : 10 grades
Scan speed : 6.25mm/s, 12.5mm/s, 25mm/s
TEMP
Accuracy : ±0.1⁰C or ±0.2⁰C ( without probe )
Measurement range : 0~50⁰C ( 32~122⁰F )
Channel : Two channels
Resolution : 0.1⁰C
Parameters : T1, T2 and TD
SpO2
Measurement range : 0-100%
Resolution : 1%
Accuracy : ±2% ( 70-100%, Adult/Paediatric); ±3% (70-100%, Neonate); 0-69%, unspecified
Refreshing Rate : 1s
Masimo SET SpO²(Optional)
Measurement range : 0-100%
Resolution : 1%
Accuracy : ±2% ( 70-100%, Adult/Pediatric, Non-Motion,
Low perfusion);
±3% ( 70-100%, Neonate, Non-motion);
0-69%, unspecified
Refreshing Rate : 1s
Pulse Rate
Range : 20~300 bpm
Resolution : 1 bpm
Accuracy : ±3bpm ( non-motion)
±5bpm ( motion)
Refreshing rate : 1s
NIBP
Measurement method : Automatic oscillometric method
Operating mode : Manual, automatic, continuous
Measurement unit : mmHg/kPa selectable
Typical measurement time : 20~40s
Measurement type : Systolic, Diastolic
Mean Measurement range (mmHg)
Range of Systolic pressure : Adult 40-270
Pediatric 40-230
Neonatal 40-135
Range of Diastolic pressure : Adult 10-210
Pediatric 10-150
Neonatal 10-100
Range of Mean pressure : Adult 20-230
Pediatric 20-165
Neonatal 20-110
Maximum average error : ±5mmHg
Maximum standard deviation : ±8mmHg
Resolution : 1mmHg
interval : 1,2,3,4,5,10,15,30,60,90,120,180,240,480 minutes
Overpressure protection : Software and hardware,
double safety protection Cuff
Pressure range : 0-280mmHg
IBP(Optional)
Channel : 2-Channel or 4-Channel
Measurement Types : Systolic, Diastolic, Mean
Range : 0mmHg~300mmHg
Resolution : 1mmHg
Accuracy : ±2% or ±1mmHg, whichever is greater(without sensor)
Sensitivity : 5uV/mmHg/V
Impedance range : 300 to 3000Ω
C.O.(Optional)
Method : Thermodilution
Range : C.O.: 0.1 to 20 L/min
TB : 23 to 43⁰C
T1 : 0 to 27⁰C
Accuracy : C.O. : ±5% or ±0.1L/min, whichever is greater
TB, T1 : ±0.1⁰C ( without sensor)
Phasein Multi-gas IRMA AX+Mainstream CO2(Optional)
Gas : C02,N2O,HAL,ISO,ENF,SEV,DES with Automatic Identification
Warm-up time : Full accuracy within 20 seconds for IRMA AX+
CO2 Accuracy : 0-10%: ±(02%+2% of the reading) 0- 15%:±(0.3% + 2% of the reading)
N2O Accurucy : 0-100%:±(2%+2% of the reading
HAL, ISO, ENF : 0-8%:±(0.15%+ 5% of the reading)
SEV : 0-10%±(0.15%+5% of the reading)
DES : 0-22%: ±(0.15% +5% of the reading)
Agent identification time : <20s(typical <10s)
AWRR range : 0-150 bpm
AWRR accuracy : +/- 1bpm
Apnea 1lme : 20s, 25s,30s,35s,40s
Recorder (Optional)
Horizontal resolution : 16 dots/mm (25 mm/s paper speed)
Vertical resolution : 8 dots/mm
Paper speed : 25 mm/s, 50 mm/s
Number of waveforms channels : 3
Phasein IRMA Sidestream CO2 (Optional)
Warm-up time : Full accuracy within 10 seconds
Sampling flow rate : 50ml/min(±10/min)
Accuracy : ±(0.2% + 2% of the reading)
Measurement Range : 0-15%
Rise time : 200ms, typical at 50ml/min flow rat
Total response time : within 3 seconds ( with 2 m Momoline sampling Line)
AWRR Range : 0-150bpm
AWRR Accuracy : ±1 breath
Operation Environment
Power : AC - 250, 50/60Hz
Temperature : 0-40⁰C
Humidity : 15-85%
Patient Range : Adult, Pediatric, Neonate