Oxi-100

Contact Us

Technical Data Sheet:


Pulse Oximeter :
SpO2 Measuring range : 0 ...100% Spo 1% Grad.
Display range : 0 ... 99% Sp0 . 1% Grad.
Accuracy : Adults +/- 2% ( 70 - 100%) +/- 1% ( 40 - 70%)
: Children +/- 3% ( 70 - 100%) +/- 2% (40 - 70%) (not specified below 70%)
Pulse rate Measuring range Accuracy : 25 ... 250 bpm with 1 Grad. : +/- 2 bpm or +/- 2% whichever is greater.
User defined Alarms Alerts : High/Low SpO and PR. 2 : Sensor Disconnect, Sensor Failure and Low Battery.
AwRR : Range - 0~150 rpm Accuracy : +/- 1 rpm(0~70rpm) Undefined within other ranges Resolution : 1 rpm
NIBP :
Measurement method : Automatic oscillometric method
Operating mode : Manual, automatic, continuous
Measurement unit : mmHg/kPa selectable
Typical measurement time : <120s
Measurement type  : Systolic, Diastolic
Mean Measurement range (mmHg)
 Range of Systolic pressure : Adult 40-270 mmHg
Pediatric : 40-200 mmHg
Range of Diastolic pressure : Adult 10-210 mmHg
Pediatric : 10-162 mmHg
Range of Mean pressure : Adult 20-230 mmHg
Pediatric : 20-175 mmHg
Measurement accuracy
 Maximum average error  : ±5mmHg
Maximum standard deviation : ±8mmHg
Resolution : 1mmHg
interval : 1,2,3,4,5,10,15,30,60,90,120,180 minutes
Overpressure protection : Software and hardware, double safety protection Cuff
Pressure range : 0-300mmHg
Alarm Ranges 
Range of Systolic pressure : Adult 40-270 mmHg
Pediatric  : 40-200 mmHg
Range of Diastolic pressure : Adult 10-210 mmHg
Pediatric : 10-162 mmHg
Range of Mean pressure : Adult 20-230 mmHg
Pediatric : 20-175 mmHg
TEMPERATURE :
Range  : 0 - 50 C
Accuracy  : /- 0.1 C
Resolution  : 0.1
eTCO2:
METHOD : Infrared radiation absorption technology
CO2 MEASUREMENT RANGE  : 0 ~ 20 Vol%
ACCURACY  : 0 ~ 12 %: ± (0.2 Vol% + 2% of reading),12 ~ 20 % : ± (0.2 Vol% + 6% of reading)
MEASUREMENT ACCURACY DRIFT : accuracy requirements within 6 hours
RESOLUTION : 0.1 Vol%
ACCURACY :  Mean error: <±5mmHg, Standard deviation: <8mmHg
APNEA ALARM DELAY TIME : 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s
ALARM RANGE : EtCO2: 0 ~ 150mmHg, FiCO2: 0 ~ 150mmHg, awRR: 0 ~ 150rpm
HMI :
Display : 5" Colour TFT LCD, 800 x 480 pixels
Audio/Visual indicators  : Alarm Limit reached, Alarm tone, Alarm mute, pulse strength, Patient type, time, battery status, connection status
LEDs : Adult, Neonate, Pressure Unit or SpO2/PR (according to model), battery in use, battery charging,silenced alarm.
User interface Languages  : English
Physical And Electrical Parameters :
Mains : 115/230 V, 50/60 Hz
Battery  : NiMH, approx. 6 h operation
Power consumption  : 1,4 W
Dimensions  : 230 mm x 200 mm x 80 m
Weight  : 1.650 g
Housing : ABS