Oxi-100
Technical Data Sheet: |
||
Pulse Oximeter : | ||
SpO2 Measuring range | : | 0 ...100% Spo 1% Grad. |
Display range | : | 0 ... 99% Sp0 . 1% Grad. |
Accuracy | : | Adults +/- 2% ( 70 - 100%) +/- 1% ( 40 - 70%) |
: | Children +/- 3% ( 70 - 100%) +/- 2% (40 - 70%) (not specified below 70%) | |
Pulse rate Measuring range Accuracy | : | 25 ... 250 bpm with 1 Grad. : +/- 2 bpm or +/- 2% whichever is greater. |
User defined Alarms Alerts | : | High/Low SpO and PR. 2 : Sensor Disconnect, Sensor Failure and Low Battery. |
AwRR | : | Range - 0~150 rpm Accuracy : +/- 1 rpm(0~70rpm) Undefined within other ranges Resolution : 1 rpm |
NIBP : | ||
Measurement method | : | Automatic oscillometric method |
Operating mode | : | Manual, automatic, continuous |
Measurement unit | : | mmHg/kPa selectable |
Typical measurement time | : | <120s |
Measurement type | : | Systolic, Diastolic |
Mean Measurement range (mmHg) | ||
Range of Systolic pressure | : | Adult 40-270 mmHg |
Pediatric | : | 40-200 mmHg |
Range of Diastolic pressure | : | Adult 10-210 mmHg |
Pediatric | : | 10-162 mmHg |
Range of Mean pressure | : | Adult 20-230 mmHg |
Pediatric | : | 20-175 mmHg |
Measurement accuracy | ||
Maximum average error | : | ±5mmHg |
Maximum standard deviation | : | ±8mmHg |
Resolution | : | 1mmHg |
interval | : | 1,2,3,4,5,10,15,30,60,90,120,180 minutes |
Overpressure protection | : | Software and hardware, double safety protection Cuff |
Pressure range | : | 0-300mmHg |
Alarm Ranges | ||
Range of Systolic pressure | : | Adult 40-270 mmHg |
Pediatric | : | 40-200 mmHg |
Range of Diastolic pressure | : | Adult 10-210 mmHg |
Pediatric | : | 10-162 mmHg |
Range of Mean pressure | : | Adult 20-230 mmHg |
Pediatric | : | 20-175 mmHg |
TEMPERATURE : | ||
Range | : | 0 - 50 C |
Accuracy | : | /- 0.1 C |
Resolution | : | 0.1 |
eTCO2: | ||
METHOD | : | Infrared radiation absorption technology |
CO2 MEASUREMENT RANGE | : | 0 ~ 20 Vol% |
ACCURACY | : | 0 ~ 12 %: ± (0.2 Vol% + 2% of reading),12 ~ 20 % : ± (0.2 Vol% + 6% of reading) |
MEASUREMENT ACCURACY DRIFT | : | accuracy requirements within 6 hours |
RESOLUTION | : | 0.1 Vol% |
ACCURACY | : | Mean error: <±5mmHg, Standard deviation: <8mmHg |
APNEA ALARM DELAY TIME | : | 20s, 25s, 30s, 35s, 40s, 45s, 50s, 55s, 60s |
ALARM RANGE | : | EtCO2: 0 ~ 150mmHg, FiCO2: 0 ~ 150mmHg, awRR: 0 ~ 150rpm |
HMI : | ||
Display | : | 5" Colour TFT LCD, 800 x 480 pixels |
Audio/Visual indicators | : | Alarm Limit reached, Alarm tone, Alarm mute, pulse strength, Patient type, time, battery status, connection status |
LEDs | : | Adult, Neonate, Pressure Unit or SpO2/PR (according to model), battery in use, battery charging,silenced alarm. |
User interface Languages | : | English |
Physical And Electrical Parameters : | ||
Mains | : | 115/230 V, 50/60 Hz |
Battery | : | NiMH, approx. 6 h operation |
Power consumption | : | 1,4 W |
Dimensions | : | 230 mm x 200 mm x 80 m |
Weight | : | 1.650 g |
Housing | : | ABS |